(주)코린스

분류 검색

현재위치 : Home > > >

 
이전 큰 이미지 다음

ML2222 Bench Calibrator Mulitifunction

제조사 : DAVIS
제품번호:ML2222
시중가격 : 원
판매가격  : 원
포인트 : 점
수량 :
  PLUS   MINUS   개
■ 회원가입을 하시면 할인된가격을 보실수 있습니다.
■ 모든가격은부가세별도 입니다.
■ 회원이 되시면 가격변동정보, 신제품, 특판제품, 기술정보등 다양한
뉴스정보를 보내드립니다.
■ 30만원미만은 택배비 3,000원(착불)입니다.
■ 카드결재시 세금계산서 발행 불가합니다.
■ 배송은 평균 2~3일정도 예상됩니다.

상세페이지를 마이피플로 보내기 - 웹용 상세페이지를 twitter로 보내기 상세페이지를 facebook으로 보내기 상세페이지를 Me2Day로 보내기

The ML2222 Bench Calibrator is a revolutionary new product design that offers laboratory grade accuracy at a handheld calibrator price.

 

The ML2222 has user friendly control functions with direct keyboard and scroll capability for data entry. It can also be operated through an RS232, or optional IEEE interface for automated testing. The ML2000 is compatible and works with Fluke Met/Cal automated calibration software.

 

The ML2222 has the ability to store up to 9 setpoints for each output range. The setpoints can be selected manually or automatically stepped at timed intervals.

 

The ML2222 contains a standby/operated safety feature typically found in calibrators selling for considerably more than the ML2222. This feature protects both the operator and the calibrator from potential overload conditions.

 

For superior calibration accuracy the ML2222 features Tellurium copper binding posts reducing thermal emf’s to less than 1 microvolt.
Features
  • Handheld and bench multimeters
  • Analog voltage and current instruments
  • Data loggers
  • Strip chart recorders
  • Other related instruments
  • Lab Standard
  • Engineers involved in product development where sensors, A/D converters, data acquisition and process control circuitry are being developed
  • Production testing of devices that need analog signals as a stimulus
  • Process control calibrator reference instrument standards (used in plant I.E. shops)
  • Cal labs that have a heavy DC calibration workload
SPECIFICATIONS

Power Requirements

100/120/240 VAC
Remote Interface RS-232 and GPIB (488 optional)
Circuit Protection Self resetting along with conventional fuses
Warm-up Time 30 min. to best accuracy
Temperature Range
Operating 0°C to 40°C
Storage -20°C to 60°C
Stability For temps less than -18°C or greater than 28°C, 10% of accuracy spec/°C
Display 16 x 2 large character alphanumeric with LED backlight
Size Approx. 11.5" x 4.7" x 12"
Weight Approx. 5 pounds
Range Accuracy (% of Reading) (1 year spec at 23°C ±5°C) Resolution Max Burden
0-100.000 mV ±0.003% ±3 μV 1 μV 10 mA
0-1.00000 V ±0.003% ±20 μV 10 μV 10 mA
0-10.0000 V ±0.003% ±200 μV 100 μV 10 mA
0-100.000 V ±0.003% ±2 mV 1 mV 1 mA
0-100.000 mA ±0.01% ±2 μA 1 μA 10 V

CAT NO. MODEL DEscRIPTION AVAIL PRICE
ML2222 M2000 Voltage/Current Bench Calibrator Yes 3,112,020