The TRX allows for fast and easy operation with its user-friendly design. By scrolling through the menu, the display will prompt the operator through calibrations by asking questions such as : Select ¡°INPUT or OUTPUT,¡± select MILLIAMPS, MILLIVOLTS, T/C¡±, etc. Settings can be changed at any time by pressing the ¡°C/CE¡± key which allows you to step backwards in the program and after previous settings.
The programmable step function allows for storage of up to six calibration values. These values can be recalled manually in sequence or be sent out automatically over a preprogrammed time interval. The ¡°FIXED STEPS¡± key allows for output of signals in 0%, 25%, 50%, 75% and 100% of selected span value.
The TRX offers a wide choice of measuring, simulation and combination possibilities including: Steps, Ramping, Temp Xmtr Calibration, Scaling, Selectable Standards, Cold Junction Compensation, Loop Testing, Speed Setting, etc.
The TRX will allow you to measure and simulate millivolts, Volts, Ohms, RTD¡¯s, Thermocouples, Milliamps, and also provide a 24 VDC power supply for 2/3 wire 4-20 mA or 0-20 mA transmitters.
The TRX is the ideal tool for the instrument engineer as it can act as a simulator, receiving controller, power supply and decade box to make maintenance, checking and calibration of many devices including RTD¡¯s and T/C¡¯s very easy.
Features
All in one instrument
Cold junction compensated
Measures signals
Simulates signals
Rugged filed instrument
Compact size
Self testing
Programmable
Ramp functions
Complete XMTR calibration
Step functions
mA scaling
IEC/ANSI-DIN-JIS curves
Optional: NIST traceable certificate
Instrinsically safe model available
SPECIFICATIONS
Reference
20¡ÆC (68¡ÆF)¡¾3K
Long Term Drift
¡¾0.03% of range/year
Indicated Accuracies
Specified for 15¡ÆC (60¡ÆF) to 35¡ÆC (95¡ÆF).
Outside these limits: ¡¾1 digit on zero and ¡¾0.005% (0.0025%) of range/¡ÆC(¡ÆF)
Operating Temperature
-10¡ÆC to +50¡ÆC (14¡ÆF to 122¡ÆF)
Storage Temperature
-20¡ÆC to +70¡ÆC (-4¡ÆF to 158¡ÆF)
Calibration
Traceable to National Standards with correlation to NBS
Warm Up Time
2 minutes in a constant ambient temperature
Relative Humidity
0 to 90% non-condensing
Cold Junction Compensation
¡¾0.25¡ÆC (0.5¡ÆF) with the Pt100 sensor in the connection module or 0.8¡ÆC (1.4¡ÆF) with the semi-conductor sensor in model MLC housing or to be set by the user with 99¡ÆC/¡ÆF to 99¡ÆC/¡ÆF
Excitation Current, RTD Simulation
5mA max,either polarity (from an external source)
Read-out
41/2 or 4 digits, depending on the selected function, Text in English
Batteries
4 x 1.5 Volt, type LR14 or size C
Battery Life
25 hours with alkaline batteries at 20¡ÆC (68¡ÆF), 12 hours with 20 mA load
Low Battery Indication
Pre-warning alternately flashes "poor battery" after approx. 15 minutes the unit stops and indicates steady "replace batteries"
External Power Supply
2.5mm plug, 6 Volt at 300mA
Connections
Suitable for 2mm banana plugs or loose wire end (with connection module)
Protection
IP53
Housing
Textured high-impact ABS plastic
Hazardous Area
Designed for use in div. 2 classified areas
Size
200 x 117 x 32mm (8.0 x 4.7 x 1.3") without carrying case
Weight
0.7kg (1.65 pound) including batteries and carrying case
SPECIFICATIONS
Function
Range
Range
Resolution
Accuracy
Remarks
Measure mVolt
0-120mV
0.01mV
¡¾0.025%
R-input>20M¥Ø
Output mVolt
-10-120mV
0.01mV
¡¾0.025%
R-output 0,2¥Ø
Measure Volt
0-120V
0.01V
¡¾0.05%
R-input>1M¥Ø
Output Volt
0-12V
0.001V
¡¾0.025%
R-output 0.2¥Ø
Measure mA
0-52mA
0.01mA
¡¾0.05%
R-input 10¥Ø
Output mA
0-24mA
0.01mA
¡¾0.05%
R-max. 900¥Ø
Transmitter Sim.
0-24mA
0.01mA
¡¾0.05%
V-max ext. 56 volts
Meas./Sim. Ohm
0-390¥Ø
0.1¥Ø
¡¾0.1¥Ø
For RTD's and Ohms ¡¾0.25¡ÆC/0.5¡ÆF 1.25mA excitation current, Curves IEC 751 (385) or JIS (392)
Thermocouples J,K,T,B,R,S and E in accordance with IEC standard 584-1 or DIN 43710 for U and L
Meas./Sim. Pt100
-200/850¡ÆC
-328/1562¡ÆF
0.25¡ÆC/0.5¡ÆF
¡¾0.25¡ÆC/0.5¡ÆF
Meas./Sim. Nl100
-60/850¡ÆC
-76/482¡ÆF
0.25¡ÆC/0.5¡ÆF
¡¾0.25¡ÆC/0.5¡ÆF
Meas./Sim. TC-J/L
-210/1190¡ÆC
-346/2174¡ÆF
0.2¡ÆC/0.4¡ÆF
¡¾0.4¡ÆC/0.8¡ÆF
Meas./Sim. TC-K
-210/1370¡ÆC
-346/2498¡ÆF
0.2¡ÆC/0.4¡ÆF
¡¾0.6¡ÆC/1.2¡ÆF
Meas./Sim. TC-K
-230/-210¡ÆC
-382/-346¡ÆF
1¡ÆC/2¡ÆF
¡¾2¡ÆC/4¡ÆF
Meas./Sim. TC-T/U
-75/400¡ÆC
-103/752¡ÆF
0.2¡ÆC/0.4¡ÆF
¡¾0.5¡ÆC/1¡ÆF
Meas./Sim. TC-T/U
-180/-75¡ÆC
-292/-103¡ÆF
0.4¡ÆC/0.8¡ÆF
¡¾1¡ÆC/2¡ÆF
Meas./Sim. TC-T/U
-250/-180¡ÆC
-418/-292¡ÆF
0.6¡ÆC/1.2¡ÆF
¡¾1.6¡ÆC/3.2¡ÆF
Meas./Sim.TC-B
540/1810¡ÆC
1004/3290¡ÆC
1¡ÆC/2¡ÆF
¡¾2.5¡ÆC/5¡ÆF
Meas./Sim. TC-B
315/540¡ÆC
599/1004¡ÆF
2¡ÆC/4¡ÆF
¡¾6¡ÆC/12¡ÆF
Meas./Sim. TC-B
180/315¡ÆC
356/599¡ÆF
4¡ÆC/7¡ÆF
¡¾8¡ÆC/16¡ÆF
Meas./Sim TC-R
-50/1760¡ÆC
-58/3200¡ÆF
0.7¡ÆC/1.4¡ÆF
¡¾2¡ÆC/4¡ÆF
Meas./Sim. TC-S
-50/1760¡ÆC
-58/3200¡ÆF
0.8¡ÆC/1.6¡ÆF
¡¾2¡ÆC/4¡ÆF
Meas./Sim. TC-E
-240/1000¡ÆC
-400/1832¡ÆF
0.1¡ÆC/0.2¡ÆF
¡¾0.3¡ÆC/0.6¡ÆF
Meas./Sim TC-E
-250/-240¡ÆC
-418/-400¡ÆF
0.8¡ÆC/1.6¡ÆF
¡¾2¡ÆC/4¡ÆF
Meas./Sim TC-N
0/1300¡ÆC
32/2372¡ÆF
0.2¡ÆC/0.4¡ÆF
¡¾0.6¡ÆC/1.2¡ÆF
Notes: Indicated ranges for thermocouples are in accordance with IEC 584-1 Menu selectable curves. IEC/ANSI DIN or JIS
CAT NO.
MODEL
DEscRIPTION
AVAIL
PRICE
JF150310
TRX
Portable Signal Calibrator
Yes
3,662,820
With carrying case, set of batteries, instruction guide spare fuse, terminal block and test leads
UM13603
191-A005
OPTIONAL Power Supply, 70V FOR Portable Signal Calibrator
Yes
201,960
UM050
191-A006
OPTIONAL Terminal Block FOR Portable Signal Calibrator
Yes
358,020
UM500
191-A014
OPTIONAL Extension Cable Kit FOR Portable Signal Calibrator