No.7210-00/7215-00 SIGMA II THERMOHYGROGRAPH
SPECIFICATIONS

Cat. No.:

No.7210-00 (Quartz type),
No.7215-00 (Spring-wound type)
Model: NSII-Q (Quartz type), NSII-S (Spring-wound type)
Clock: Quartz clock: 1-day, 7-day and 32-day by change-over switch
Battery: 2 pcs, of "C" cell (1.5 V)
Life of battery: 1 year for continuous use
Spring-wound: 1-day and 7-day changed by gear
Sensors: Aged bimetalic strip for temperature
Human hair bundle for relative humidity
Measuring range: -15 to+40oC (option: -10 to 50oC) for temperature
0 to 100% for relative humidity
Accuracy: 1oC for temperature 3% for relative humidity
Chart graduation: 1oC for temperature 1% for relative humidity
Recording pens: Recording pens: Cartridge pens (violet)
Dimensions: 336 (W)x148 (D)x295 (H) mm
Weight: Approx. 3.1 kg (Quartz type)
Approx. 3.5 kg (Spring-wound type)
Accessories: 7-day chart for 1 year use (55 sheets/pad)
chart holder
Instruction manual, "C" cell x 2 pcs.
(applicable to quartz type only)
Accessories Consumable
No. 7210-60: 1-day chart (400 sheets/pad)
No. 7210-62: 7-day chart (55 sheets/pad): Dustproof type
No. 7210-64: 32-day chart (55 sheets/pad): Dustproof type
No. 7210-90: Spare cartridge pen (violet)
No. 7210-92: Spare cartridge pen (red)
No. 7210-20: Axial AC fan

No.7210-20 Axial AC Fan for SIGMA II Series Reorders
Use an external fan for faster response and more accurate measurement. It is in combination with this fan that the maximum capability of measurement for temperature and humidity can be achieved.
The fan inhales fresh air into the instrument and makes the response of sensor faster, resulting in the detection of accurate data. It is effective, particularly in warehouses which are excluded from the open air, laboratories where ventilation is poor or air tight small space.

¡¡

Cat. No. 7210-20
Model Axial AC Fan
Power requirements 100 to 120 VAC 50/60 Hz
Power consumption 70/55 mA
Revolution of fan 2700/3300 rpm
Max. airflow 0.73/0.94 m3 min
Max. static pressure 3.4/5 mmAq