The ML2222 Bench Calibrator is a revolutionary new product design that offers laboratory grade accuracy at a handheld calibrator price.
¡¡
The ML2222 has user friendly control functions with direct keyboard and scroll capability for data entry. It can also be operated through an RS232, or optional IEEE interface for automated testing. The ML2000 is compatible and works with Fluke Met/Cal automated calibration software.
¡¡
The ML2222 has the ability to store up to 9 setpoints for each output range. The setpoints can be selected manually or automatically stepped at timed intervals.
¡¡
The ML2222 contains a standby/operated safety feature typically found in calibrators selling for considerably more than the ML2222. This feature protects both the operator and the calibrator from potential overload conditions.
¡¡
For superior calibration accuracy the ML2222 features Tellurium copper binding posts reducing thermal emf¡¯s to less than 1 microvolt.
Features
Handheld and bench multimeters
Analog voltage and current instruments
Data loggers
Strip chart recorders
Other related instruments
Lab Standard
Engineers involved in product development where sensors, A/D converters, data acquisition and process control circuitry are being developed
Production testing of devices that need analog signals as a stimulus
Process control calibrator reference instrument standards (used in plant I.E. shops)
Cal labs that have a heavy DC calibration workload
SPECIFICATIONS
Power Requirements
100/120/240 VAC
Remote Interface
RS-232 and GPIB (488 optional)
Circuit Protection
Self resetting along with conventional fuses
Warm-up Time
30 min. to best accuracy
Temperature Range
Operating
0¡ÆC to 40¡ÆC
Storage
-20¡ÆC to 60¡ÆC
Stability
For temps less than -18¡ÆC or greater than 28¡ÆC, 10% of accuracy spec/¡ÆC
Display
16 x 2 large character alphanumeric with LED backlight
Size
Approx. 11.5" x 4.7" x 12"
Weight
Approx. 5 pounds
Range
Accuracy (% of Reading) (1 year spec at 23¡ÆC ¡¾5¡ÆC)