Tactile keypad provides readout in H2O, Hg, PSI, kPa, mmHg, mBAR, cmH2O, and BARS. |
Simple to Use Tactile keypad provides readout in H2O, Hg, PSI, kPa, mmHg, mBAR, cmH2O, and BARS. One user-defined scale can also be programmmed. Temperature scaling for H2O measurements is selectable among 4¡ÆC, 20¡ÆC, and 60¡ÆF.
Test Pressure Switches Min/max memory reads out the highest and lowest measurements. Trip Detect monitors the electrical measurement for a change of state, and freezes the last pressure measurement on the display. The unit can also display input vs. output as a percent error. A damping function that averages the last 16 readings to compensate for pulsating pressures is built in.
Each unit includes: Set of test leads, internal NiCd battery pack, battery charger, carrying case, manufacturer¡¯s NIST traceable certificate, and instruction manual.
This unit is available in a CSA intrinsically safe version for Class 1 Groups A, B, C, and D. |
|
Features |
- Choice of dry (non-isolated) or wet (isolated) sensors
- Intrinsically safe option (see ordering Information)
- Single or dual sensor modes
- Also reads mA and Volts
- 0.1% or better accuracy
- Five (5) digit resolution
|
| | SPECIFICATIONS – Range Selection Chart
Range |
Range Code |
Gauge |
Dry |
Wet |
0.3 PSIG |
01 |
NA |
1 |
02 |
NA |
5 |
03 |
60 |
7.2 |
22 |
61 |
10 |
NA |
62 |
15 |
24 |
63 |
30 |
04 |
64 |
50 |
05 |
65 |
100 |
06 |
66 |
150 |
NA |
67 |
300 |
07 |
68 |
1,000 |
08 |
08 |
1,500 |
23 |
23 |
3,000 |
09 |
09 |
5,000 |
10 |
10 |
Vacuum |
-1 PSI |
11 |
NA |
-5 |
12 |
NA |
-15 |
13 |
NA |
Absolute |
15 PSIA |
14 |
50 |
30 |
15 |
51 |
50 |
16 |
52 |
100 |
NA |
53 |
Differential |
1 PSID |
17 |
NA |
5 |
18 |
NA |
30 |
19 |
NA |
50 |
20 |
NA |
Note: Dry indicates the sensor is approved for use with dry media (gasses) only. Wet indicates the sensor is approved for use with any media compatible with 316 SS. |
Accuracy |
0.07 % rdg + 0.03% of scale + 1 digit or better, all PSI ranges |
Resolution |
50,000 counts max. depending on range and engineering units |
mA Input |
0-20 mA and 0-50 mA ranges |
V Input |
0-30 VDC range |
mA/V Accuracy |
0.025% ¡¾1 LSD |
Media Compatibility |
Non-corrosive liquids and gasses; liquids not recommended for low pressure (0.3-100 PSIG) ranges |
Pressure Fittings |
1/8" FNTP 316 SS |
Display |
Two-line, 32 character alphanumeric LCD |
Power Supply |
Four AA rechargeable NiCd batteries |
CAT NO. |
|
MODEL |
|
DEscRIPTION |
|
AVAIL |
|
PRICE |
BE3201 |
|
320-*-70-IS |
|
Single Sensor, 70VAC Charger, no loop supply, intrinsically safe |
|
Yes |
|
3,769,200 |
Different models are single range only. |
BE3202 |
|
320-*-220-01-IS |
|
Single Sensor, 220VAC Charger, no loop supply, intrinsically safe |
|
Yes |
|
3,769,200 |
Note: For Single sensor Units, "Left Port" selection is mandatory. For Dual Sensor units, "Left Port" must be the lower of the two ranegs selected. |
BE3203 |
|
320-*-70-02 |
|
Single Sensor, 70VAC charger, loop supply |
|
Yes |
|
3,823,200 |
Note: Single Sensor units, "Left Port" selection is mandatory. For Dual Sensor units, "Left Port" must be the lower of the two ranges selected. Difffernt models are single range only. |
BE3204 |
|
320-*-220-02 |
|
Single Sensor, 220VAC Charger, loop supply |
|
Yes |
|
3,823,200 |
Note: For Single Sensor units, "Left Port" selection is mandatory. For Dual Sensors units, "Left Port" must be the lower of the two ranges selected. Different models are single range only. |
BE3205 |
|
320-*-70-01 |
|
Single Sensor, 70VAC charger |
|
Yes |
|
3,013,200 |
Note: For Single Sensor units, "Left Port" selection is mandatory. For Dual Sensor units, "Left Port" must be the lower of the two ranges selected. Different models are single range only. |
BE3206 |
|
320-*-220-01 |
|
Single Senso, 220VAC charger |
|
Yes |
|
3,013,200 |
Note: For Single Sensor units, "Left Port" is mandatory. For Dual Sensor units, "Left Port" must be the lower if the two ranges selected. Different models are single range only. |
BE3207 |
|
320-*-*-70-01-IS |
|
Dual Sensor, 70VAC Charger, no loop supply, intrinsically safe |
|
Yes |
|
5,065,200 |
Note: For Single Sensor units, "Left Port" is mandatory. For Dual Sensor units, "Left Port" must be the lower of the two ranges selected. Different models are single range only. |
BE3208 |
|
320-*-*-220-01-IS |
|
Dual Sensor, 220VAC Charger, no loop supply, intrinsically safe |
|
Yes |
|
5,065,200 |
Note: For Single Sensor units, "Left Port" is mandatory. For Dual Sensor units, "Left Port" must be the lower of the two ranges selected. Different models are single range only. |
BE3209 |
|
320-*-*-70-02 |
|
Dual Sensor, 70VAC charger, loop supply |
|
Yes |
|
5,119,200 |
Note: For Single Sensor units, "Left Port" is mandatory. For Dual Sensor units, "Left Port" must be the lower of the twp ranges selected. Different models are single range only. |
BE3210 |
|
320-*-*-220-02 |
|
Dual Sensor, 220VAC Charger, loop supply |
|
Yes |
|
5,119,200 |
Note:For Single Sensor units, "Left Port" is mandatory. For Dual Sensor units, "Left Port" must be the lowest of the two ranges selected. Different models are single range only. |
BE3211 |
|
320-*-*-70-01 |
|
Dual Sensor, 70VAC charger |
|
Yes |
|
4,309,200 |
Note: For Single Sensor units, "Left Port" is mandatory. For Dual Sensor units, "Left Port" must be the lower of the two ranges selected. Differnet models are single range only. |
BE3212 |
|
320-*-*-220-01 |
|
Dual Sensor, 220VAC charger |
|
Yes |
|
4,309,200 |
Note: For Single Sensor units, "Left Port" is mandatory. For Dual Sensor units, "Left Port" must be the lower of the two ranges selected. Different models are single range only. |
| |
|
|
| |
|