The unit is an extremely versatile and cost-effective multifunction process calibrator, which replaces many single function instruments with a single compact handheld unit. The unit measures, sources and simulates most process signals and displays both input/output readings simultaneously. |
|
Features |
- Measure/source: mA, mV, V
- Measure/simulate: 10 thermocouples, 2 RTD¡¯s, ohms.
- Dual display: Any input value with any output value
- Accuracy: 0.04% rdg for mA, mV, V, ohms
- Input/output display: 2 x 16 character dot matrix LCD
- Electrical connectors: 4mm separated input/output connectors.
- Cold junction: Automatic or manual compensation for thermocouple cold junction temperature
- Scaled readings: Process signals can be scaled into engineering units e.g. measure mA and read in ¡ÆF or ¡ÆC.
- Step Output: Step output, up and down, using arrow keys.
- Ramp Output: Programmable end points, rate and dwell.
- Preset memory: Quick store and recall of three preset values.
- Loop power: 24V loop power for mA measure and source.
|
| | SPECIFICATIONS
Measure/source |
Range |
Accuracy (1 Year) |
Resolution |
Remark |
mV |
¡¾21mV ¡¾210nmV |
0.07% + 7* 0.04% + 4* |
0.001mV 0.01mV |
R in >1000M ohm R in >1000M ohm |
V |
¡¾2.1V |
0.04% + 4* |
0.0001V |
R in >1000M ohm |
V (input) |
¡¾21V |
0.04% + 4* |
0.001V |
R in >1M ohm |
V (output) |
¡¾10.5V |
0.04% + 4* |
0.001V |
10mA max. |
mA |
¡¾21mA |
0.04% + 4* |
0.001mA |
R load <=1000 ohm R in 10 ohm |
Ohms |
0 to 400/3000 ohm |
0.04% + 4* |
0.1/1W |
V in 60V max. 0.5mA measure |
J (1) |
-200 to 1000¡ÆC |
0.7¡ÆC |
0.1¡ÆC |
0.1 to 3mA source |
L (1) |
-200 to 900¡ÆC |
0.7¡ÆC |
0.1¡ÆC |
R in >1000M ohm |
K (1) |
-250 to 1350¡ÆC |
1¡ÆC |
0.2¡ÆC |
R in >1000M ohm |
T (1) and U (2) |
-250 to 400¡ÆC |
0.6¡ÆC |
0.1¡ÆC |
R in >1000M ohm |
E (1) |
-250 to 1000¡ÆC |
0.7¡ÆC |
0.1¡ÆC |
R in >1000M ohm |
N (1) |
-250 to 1300¡ÆC |
1.4¡ÆC |
0.5¡ÆC |
R in >1000M ohm |
R (1) and S (1) |
-50 to 1750¡ÆC |
1.4¡ÆC |
0.5¡ÆC |
R in >1000M ohm |
B (1) |
-200 to 1820¡ÆC |
2.8¡ÆC |
1¡ÆC |
R in >1000M ohm |
Pt 100 (3), (4) |
-200 to 850¡ÆC |
0.6¡ÆC |
0.1¡ÆC |
R in >1000M ohm |
Ni 100 (3) |
-60 to 180¡ÆC |
0.7¡ÆC |
0.1¡ÆC |
0.5mA measure 0.1 to 3mA source 0.5mA measure 0.1 to 3mA source |
* % of reading and number of counts. Accuracy includes temperature errors 59¡Æ to 95¡ÆF (15¡Æ to 35¡ÆC). RTD and T/C accuracy best case mid range. (1) IEC 584, (2) DIN43710, (3) DIN43760, (4) IEC 751 |
Display |
2 rows of 16 characters, dot matrix LCD, 5 readings/sec. |
Temperature units |
¡ÆF and ¡ÆC . |
Cold Junction error |
¡¾0.3¡ÆC. |
Loop Power |
24V, ¡¾3%. |
Calibration Reference |
23¡ÆC ¡¾5¡ÆC. |
Operating Temperature Range |
14¡Æ to 131¡ÆF (-10¡Æ to 55¡ÆC). |
Humidity |
10 to 90% non-condensing. |
Conformity |
EN61326-1, EN61010-1, CE marked. |
Maximum Input Voltage |
30V with respect to ground. |
Power Supply |
4 AA batteries. Alkaline cells give 20 hours use at 68¨ùF in measure mode. Auto power off, battery level indicator. |
Physical |
1.75 lbs., 9.6 x 4.9 x 2 in. |
CAT NO. |
|
MODEL |
|
DEscRIPTION |
|
AVAIL |
|
PRICE |
DR72011 |
|
DPI-720 |
|
Multifunction Process Calibrator |
|
Yes |
|
1,826,820 |
Complete instrument includes one set of four AA alkaline batteries, one set of test leads, carrying case, user guide, quick reference guide and NIST traceable calibration certificate with data |
DR72012 |
|
860-174 |
|
OPTIONAL Spare Carrying Case FOR Multifunction Process Calibrator |
|
Yes |
|
73,440
| | |
|
|
| |
|
¹øÈ£ |
Á¦¸ñ |
ÀÛ¼ºÀÚ |
ÀÛ¼ºÀÏ |
Æò°¡Á¡¼ö |
|
|
|
ÀÌ »óÇ°¿¡ ´ëÇÑ »ç¿ëÈıⰡ ¾ÆÁ÷ ¾ø½À´Ï´Ù.
»ç¿ëÈı⸦ ÀÛ¼ºÇØ ÁÖ½Ã¸é ´Ù¸¥ ºÐµé²² ¸¹Àº µµ¿òÀÌ µË´Ï´Ù. |
|
|
|
|
|
|
| |
|
¹øÈ£ |
Á¦¸ñ |
ÀÛ¼ºÀÚ |
ÀÛ¼ºÀÏ |
´äº¯ |
|
|
|
ÀÌ »óÇ°¿¡ ´ëÇÑ Áú¹®ÀÌ ¾ÆÁ÷ ¾ø½À´Ï´Ù.
±Ã±ÝÇϽŠ»çÇ×Àº ÀÌ°÷¿¡ Áú¹®ÇÏ¿© ÁֽʽÿÀ. |
|
|
|
|
|
|
| |
ÀÌ »óÇ°°ú °ü·ÃµÈ »óÇ°ÀÌ ¾ø½À´Ï´Ù. | |
|
|