Recording |
Pen Type |
Disposable fiber tip |
Pen Color |
Pen 1 - Red |
Pen 2 - Green |
Chart Size |
10 inch |
Chart Drive |
Stepper motor |
Chart Rotation |
User configurable; 8 hour, 12 hour, 24 hour or 7 day |
Chart Span |
Bottom and top of span, -9999 to 9999 units |
Inputs |
Input Types/Range |
Type |
Range |
Thermocouple |
J |
0¡ÆC to 70¡ÆC |
0¡ÆF to 1400¡ÆF |
K |
0¡ÆC to 1360¡ÆC |
0¡ÆF to 2500¡ÆF |
T |
-200¡ÆC to 400¡ÆC |
-330¡ÆF to 750¡ÆF |
R |
200¡ÆC to 1650¡ÆC |
400¡ÆF to 3000¡ÆF |
S |
200¡ÆC to 1650¡ÆC |
400¡ÆF to 3000¡ÆF |
RTD |
100 ohm |
-140¡ÆC to 400¡ÆC |
-220¡ÆF to 750¡ÆF |
Platinum |
| ¡¡ |
.00385 ohms/ohm/¡ÆC |
| ¡¡ |
Current DC |
0 to 20mA, 4 to 20mA |
Internal 4.7 ohm Shunt Resistor |
Voltage DC |
0 to 25mV, 0 to 50mV, 10 to 50mV, 0 to 5 V, 1 to 5 V |
Input Performance |
Performance Under Reference Condition |
Measurement Error |
Type J, K, T, R, S and RTD: ¡¾0.25% of span ¡¾1 degree mA, mV and VDC: ¡¾0.25% of scaled span plus 1 least significant digit |
Cold Junction Compensation Error |
¡¾0.2¡ÆC @ 25¡ÆC |
Cold Junction Compensation Rejection |
0.04¡Æ/¡ÆC deviation from 25¡ÆC |
Linearization Error expectations |
TCs: ¡¾0.25¡ÆC typical, ¡¾0.5¡ÆC worst case with RTDs: ¡¾0.1¡ÆC typical, ¡¾0.3¡ÆC worst case |
Ambient Temperature Error |
¡¾0.01% of span per ¡ÆC deviation from 25¡ÆC |
Power Requirements |
Line Voltage |
90-264 VAC, 50/60 Hz |
Power Consumption |
18 VA Maximum |
Construction |
Enclosure |
Injection molded Noryl case and cover with acrylic window |
NEMA Rating |
NEMA 3 standard |
Mounting |
Panel or wall |
Overall Dimensions |
14" wide x 14" high x 3.8" deep (355.6mm x 355.6mm x 96.5mm) |
Panel Cutout |
12.7" wide x 12.7" high (322.58mm x 322.58mm) |
Panel Depth |
2.5" (63.5mm) |
Panel Protrusion |
1.3" (33.0mm) |
Weight |
15 lbs maximum |
Alarms |
Number |
Up to two process alarms for each of two inputs |
Type |
Process high or low |
Limit Device output |
Optional high/low limits for each input with latching |
Normally open output latches open |
Red reset button included to the right of the display |
Hysteresis |
Fully adjustable, 0 to 200 units, single sided |
Security |
Alarm setpoint changes can be prohibited |
Sensor Fault Action |
Alarms work normally in "Hi" and "Lo" conditions |
Relay Outputs |
Relays |
SPDT, contacts rated 5 amps resistive at 115 VAC, 2.5amps resistive at 230 VAC, 1/8 HP at 230 VAC (singlephase), 250 VA at 115/230 |