The Model 9009 Dual-Block Calibrator lets you calibrate at hot and cold temperatures at the same time. Double your productivity or cut your calibration time in half—either way you look at it, your in-field temperature calibrations just got easier.
The Model 9009 includes two independently controlled temperature blocks. The hot block provides temperatures from 35°C to 350°C, while the cold block covers the range -15·°C to 70°C. Both blocks are controlled by a precision temperature controller.
Features
Temperatures from -15°C to 350°C in one unit
Two wells in each block for simultaneous comparison calibrations
Rugged, lightweight, water-resistant enclosure
Controlled by precision temperature controller
SPECIFICATIONS
Hot Block
Cold Block
Range
50°C to 350°C (122°F to 662°F)
-15°C to 70°C (5°F to 230°F) in 23°C ambient (-8°C with hot block at 350°C) 18°F @ 662°F
Accuracy
±0.6°C
±0.2°C
Stability
±0.05°C
±0.05°C
Well-to-Well Uniformity
±0.1°C
±0.1°C
Heating Times
10 minutes from 25°C to 350°C
15 minutes from 25°C to 70°C
Cooling Times
30 minutes from 350°C to 100°C
16 minutes from 25°C to -15°C
Stabilization Times
8 minutes
8 minutes
Well Depth
4 inches (101.6 mm)
4 inches (101.6 mm)
Removable Inserts
Two 1/4" (6.4 mm) and two 3/16" (4.8 mm) inserts included; see Ordering Information for other available inserts